The Valleylab Force EZ Electrosurgical Generator has instant response technology that provides surgeons with improved performance at lower power settings, minimizing the risk of tissue damage and neuromuscular stimulation, and decreasing the need to "turn up the generator." The Force EZ will adjust automatically, responding to tissue changes, maintaining power delivery, and minimizing drag. Instant Response technology will reduce capacitive coupling by 30-50%. This is acheieved by limiting the RMS voltage and the high frequency harmonics.
Features
Specifications
Output Waveforms
Bipolar: 470 kHz sinusoid
Monopolor Cut
Pure: 393 kHz sinusoid
Blend: 393 kHz burts of sinusoid, recurring at 27 kHz intervals. 50% duty cycle envelope.
Monopolar Coag
Desiccation:
Low 1: 240 +/- 40 kHz sinusoid recurring at 39 kHz. 8% duty cycle
Low 2: 393 kHz sinusoid
Low 3: 393 kHz sinusoid
Fulguration:
High 1: 470 +/- 40 kHz damped sinusoid bursts with a repetition frequency of 57 kHz
High 2: 470 +/- 40 kHz damped sinusoidal bursts with a repetition frequency of 30 kHz
Output power changes by less than 15% or 5 watts, whichever is greater, as the line voltage varies from 90-135 volts and 186-264 volts (into a 300 ohm load).
High Frequency Leakage
Bipolar: Less than 60 mArms
Monopolar: Less than 150 mArms
REM Contact Quality Monitoring System
Measurement Frequency: 80 kHz +/- 10 kHz
Measurement Current: Less than 10 uA
Acceptable Resistance Ranges: REM pad - 5.135 ohms, Non-REM pad - less than 20 ohms
Acceptance range is 5-135 ohms after REM PolyHessive II return electrode is applied.